Language:
English
Date:
2014-04-24
Identifier:
ISBN 978-0-7709-0577-4
Metadata:
Description:
The data acquisition system (DAS) developed aims to get the 3D breast cancer tomography with 8 current injection channels and 117 voltage measurement channels, capable of generating and measuring voltages and currents. By adopting special electrode and careful circuit layout, the initial bandwidth without calibration can reach 5 MHz. Electrical test results show that the system has a SNR greater than 67dB at 5MHz without digital enhancement method.
Rights:
This document is the collection of papers accepted for presentation at the 15th International Conference on Biomedical Applications of Electrical Impedance Tomography. Each individual paper in this collection: © 2014 by the indicated authors. Collected work: © 2014 Andy Adler and Bartłomiej Grychtol. All rights reserved.
Document Files:
Session:
Session 02. Poster Session 1